Example
tspnet.tsp.abort(testConnection)
Stops remote instrument execution on
testConnection
.
Also see
None
tspnet.tsp.abortonconnect
This attribute contains the setting for abort on connect to a TSP-enabled instrument.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
No
Restore configuration
Instrument reset
Power cycle
Configuration script
1 (enable)
Usage
tspnet.tsp.abortonconnect =
value
value
= tspnet.tsp.abortonconnect
value
•
Enable:
1
•
Disable:
0
Details
This setting determines if the instrument sends an abort message when it attempts to connect to a
TSP-enabled instrument using the
tspnet.connect()
function.
When you send the abort command on an interface, it causes any other active interface on that
instrument to close. If you do not send an abort command (or if
tspnet.tsp.abortonconnect
is
set to 0) and another interface is active, connecting to a TSP-enabled remote instrument results in a
connection. However, the instrument will not respond to subsequent reads or executes because
control of the instrument is not obtained until an abort command has been sent.
Example
tspnet.tsp.abortonconnect = 0
Configure the instrument so that it does not
send an abort command when connecting to
a TSP-enabled instrument.
Also see