PA-855 Rev. B / 1-05

 

Model 2306NV  Battery/Charger Simulator

 

User’s Guide

 

Keithley Instruments, Inc.

 

28775 Aurora Road
Cleveland, Ohio 44139
(440) 248-0400
Fax: (440) 248-6168
www.keithley.com

 

NOTE

 

This information is to be used in conjunction with the Model 2306 Instruction manual.  The 
Instruction manual is on the Product Information CD-ROM that ships with the Model 2306NV.  
Make sure to thoroughly understand the operation of the Model 2306NV before operating or con-
necting the unit.

 

Overview

 

This User’s Guide contains operational differences between the Model 2306NV and the standard Model 2306. The model num-
ber displayed during the power up sequence is 2306NV instead of 2306. The REVISION NUMBER under MAIN MENU dis-
plays ‘NV’ between the two revision numbers as shown below:

 

REVISION NUMBER
B13   NV B01

 

Additionally, sending the identification query ‘

 

*IDN?

 

’ returns the 2306NV model number as shown below:

 

KEITHLEY INSTRUMENTS INC., MODEL 2306NV, xxxxxxx, yyyyy/zzzzz

 

Where: 

 

xxxxxxx

 

 is the serial number.

 

yyyyy

 

 is the firmware revision level of the digital board ROM. 

 

zzzzz

 

 is the firmware revision level display board ROM.

 

2306NV operation

 

Essentially, the Model 2306NV functions the same as the Model 2306 with one exception: the Model 2306 can perform DVM
measurements while the 2306NV (no voltmeter) cannot. With this in mind, note the following differences:

• The DVM menu item (DVM INPUT) has been removed. Also, the quick disconnect connectors contain non-functional

DVM + and DVM - terminals.

• The commands contained in Table 1 have changed. Sending these unsupported commands will cause errors.

 

Table 1

 

2306NV Command differences

 

Type

Command

Change

 

Independent voltage 
measurements (DVM)

:SENSe[1]:FUNCtion <name>
:SENSe2:FUNCtion <name>

Command <name> of 'DVM' and 'DVMeter' 
removed. Also, the portion of associated commands 
such as :NPLCycles <n> or :AVERage <NRf> with 
respect to DVMeter, have also been removed.

Signal Oriented

:MEASure[1][:<function>]?
:MEASure[1]:ARRay[:<function>]?
:MEASure2[:<function>]?
:MEASure2:ARRay[:<function>]? 

Command <function> 'DVM' and 'DVMeter' 
removed.

Common Commands

*IDN?

This command returns 2306NV for model type.