Figure 1-2
Connections for voltage verification tests
2.
Select the multimeter DC volts measuring function, and enable auto-ranging.
3.
Make sure the Model 2303 output is turned on.
4.
Verify output voltage accuracy for each of the voltages listed in Table 1-2. For each test
point:
• Set the Model 2303 output voltage to the indicated value. When setting the voltage,
set the compliance current to 3A.
• Allow the reading to settle.
• Verify that the multimeter reading is within the limits given in the table.
Table 1-2
Output voltage accuracy limits
Model 2303 output
voltage setting
Output voltage limits
(1 Year, 18°C-28°C)
5.00V
4.9875 to 5.0125V
10.00V
9.985 to 10.015V
15.00V
14.9825 to 15.0175V
NEXT
PREV
POWER
DISPLAY
2001 MULTIMETER
DCV
ACV
DCI
ACI
Ω
2
Ω
4
FREQ
TEMP
REL
TRIG
INFO
LOCAL
EXIT
ENTER
RANGE
RANGE
!
F
R
500V
PEAK
FRONT/REAR
2A 250V
AMPS
CAL
ST ORE RECALL
CHAN
SCAN
FILTER
MATH
CONFIGMENU
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1100V
PEAK
AUTO
W ARNING:
NO INTERNAL OPERAT OR SERVICABLE PAR TS,SERVICE BY QU ALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERAT OR SERVICABLE PARTS,SER VICE BY QU ALIFIED PERSONNEL ONL Y.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RA TING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
LINE RATING
100-240V AC
50, 60 HZ
185VA MAX
LINE FUSE
SLOWBLO W
2.5A, 250V
IEEE-488
(CHANGE IEEE ADDRESS
WITH FR ONT PANEL MENU)
REMO TE
DISPLAY
OPTION
ISOLATION FROM EARTH:
22 VOL TS MAX.
+
+
+
_
_
_
_
+
SOURCE
SENSE
D VM
IN
SOURCE
OUTPUT
0-20V, 0-5A
Model 2001 DMM
Model 2303
Input LO
Input HI
Source -
深圳德标仪器
135-1095-0799
Summary of Contents for 2303
Page 12: ...1 Performance Verification 135 1095 0799...
Page 26: ...2 Calibration 135 1095 0799...
Page 48: ...3 Disassembly 135 1095 0799...
Page 53: ...3 6 Disassembly 135 1095 0799...
Page 76: ...A Specifications 135 1095 0799...
Page 81: ...A 7 Specifications 135 1095 0799...
Page 82: ...B CalibrationReference 135 1095 0799...
Page 90: ...C CalibrationPrograms 135 1095 0799...