Calibration 14-9
Figure 14-3
Connections for 5mA range calibration
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
LINE RATING
100-120VAC, 200-240VAC
50, 60 HZ 165VA MAX
LINE FUSE
SLOWBLOW
2.0A, 250V
IEEE-488
(ENTER IEEE ADDRESS
FROM FRONT PANEL MENU)
REMOTE
DISPLAY
OPTION
+
+
+
_
_
_
_
+
SOURCE
SENSE
DVM
IN
SOURCE
OUTPUT #1
ISOLATION FROM EARTH: 22 VOLTS MAX.
RELAY
CONTROL
24VDC MAX
CAT
I
DVM IN
+30 VDC MAX
NEXT
PREV
POWER
DISPLAY
2001 MULTIMETER
DCV
ACV
DCI
ACI
Ω
2
Ω
4
FREQ
TEMP
REL
TRIG
INFO
LOCAL
EXIT
ENTER
RANGE
RANGE
!
F
R
500V
PEAK
FRONT/REAR
2A 250V
AMPS
CAL
STORE RECALL
CHAN
SCAN
FILTER MATH
CONFIG MENU
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1100V
PEAK
AUTO
Model 2001 DMM
Model 2306
Input LO
Input HI
Source -
3k
Ω
Resistor
Note: Use 4-wire connections
to resistor terminals.
Sense -
Sense +
OUTPUT #1 connections
shown. Use OUTPUT #2
for Charger Channel.
+
+
+
_
_
_
_
+
SOURCE
SENSE
DVM
IN
SOURCE
OUTPUT #2
Step 3: Perform Channel #2 (charger channel) calibration steps
1.
Connect the OUTPUT #2 SOURCE, SENSE, and DVM IN terminals to the DMM. Use
the same general connections shown in
except make all connections to the
OUTPUT #2 terminals.
DMM input connections:
HI: , SENSE +, and DVM IN +
LO: SOURCE -, SENSE -, and DVM IN -
2.
The Model 2306 will prompt you to set the full-scale Channel #2 output voltage:
CAL VOLTS CHAN 2
SET 14.0000 V
3.
Use the edit keys to set the voltage to 14.0000V, then press ENTER.
4.
The unit will prompt you for the DMM reading, which will be used to calibrate the
Channel #2 full-scale output voltage:
CAL VOLTS CHAN 2
READ1 14.0000 V
5.
Using the edit keys, adjust the Model 2306 voltage display to agree with the DMM volt-
age reading, then press ENTER. The unit will prompt for another DMM reading, which
will be used to calibrate the full-scale Channel #1 measurement function:
CAL VOLTS CHAN 2
READ2 14.0000 V
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Summary of Contents for 2302
Page 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...