Section 7: SCPI command reference
Series 2280 Precision DC Power Supply Reference Manual
7-104
077085503 / March 2019
:STATus:OPERation:INSTrument:ISUMmary:ENABle
This command sets or reads the contents of the Operation Instrument Summary Enable Register of the status
model.
Type
Affected by
Where saved
Default value
Command and query
Power cycle
Not applicable
0
Usage
:STATus:OPERation:INSTrument:ISUMmary:ENABle <n>
:STATus:OPERation:INSTrument:ISUMmary:ENABle?
[n]
The status of the Operation Instrument Summary Enable Register
Details
This command sets or reads the contents of the enable register of the Operation Instrument
Summary Event Register (OISER).
When one of these bits is set and the corresponding bit in the OISER is set, bit 1 in the Operation
Instrument Event Register is set.
When sending binary values, preface
<n>
with
#b
. When sending hexadecimal values, preface
<n>
with
#h
. No preface is needed when sending decimal values.
Example
:STAT:OPER:INST:ISUM:ENAB #b0000000100010000
Sets the filter and measure bits of the
Operation Instrument Summary Enable
Register using a binary value.
You could also send the decimal value
272
or the hexadecimal value
#h110
Also see
Operation Instrument summary Event Register
(on page D-13)