Section 7: SCPI command reference
Series 2280 Precision DC Power Supply Reference Manual
7-64
077085503 / March 2019
:SENSe[n]:<function>:AVERage:WINDow
This command specifies a filter window to control filter threshold.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
1
Usage
:SENSe[n]:<function>:AVERage:WINDow <value>
:SENSe[n]:<function>:AVERage:WINDow?
:SENSe[n]:<function>:AVERage:WINDow? DEFault
:SENSe[n]:<function>:AVERage:WINDow? MAXimum
:SENSe[n]:<function>:AVERage:WINDow? MINimum
[n]
Channel number: For single-channel instruments, the value is 1
<function>
The measurement function to which this setting applies:
Concurrent:
CONCurrent[:DC]
Current:
CURRent[:DC]
Voltage:
VOLTage[:DC]
<value>
The filter window:
0.01
to
100
. This value is expressed in terms of a percentage.
There are five acceptable window percentages: 0.01, 0.1, 1, 10, and 100.
Details
The <
value
> parameter can be set to any value from 0.01 to 100. The instrument rounds up the
value to the nearest window value. For example, the filter window is set to 1 if you send the following
command:
SENS:CONC:AVER:WIND 0.5
Example
SENS:CONC:AVER:WIND 0.1
Set the average filter window as 0.1%.
Also see
:SENSe[n]:<function>:AVERage:COUNt
(on page 7-61)
:SENSe[n]:<function>:AVERage[:STATe]
(on page 7-62)
:SENSe[n]:<function>:AVERage:TCONtrol
(on page 7-63)