Series 2230G Triple-Channel Power Supplies User's Manual
Section 5: SCPI command reference
2230G-900-01 Rev. A / June 2018
5-15
CALibrate:STRing
This command writes or queries the calibration information of the instrument. The query version of this command
returns the saved calibration information for the instrument.
Group
Calibration
Syntax
CALibrate:STRing <string>
CALibrate:STRing?
Arguments
<string>
The maximum length of the string is 22 bytes.
Default value
Not applicable
Returns
A string with a length of up to 22 bytes.
Example
CAL:STR "2017-03-28"
Sets the calibration information of the instrument to
2017-03-28
.
See also
None
CALibrate:VOLTage[:DATA]
This command sets the actual output voltage value of the calibration point.
Group
Calibration
Syntax
CALibrate:VOLTage[:DATA] <NR2>
Arguments
<NR2>
The voltage value of the calibration point.
Default value
Not applicable
Returns
Not applicable