Series 2230G Triple-Channel Power Supplies User's Manual
Section 3: Function and features
2230G-900-01 Rev. A / June 2018
3-5
Menu items
Description
Combine
Combine Off (default)
Disables the series or parallel operation mode.
V1+V2 Series
Sets CH1 and CH2 to series operation mode.
I1+I2 Parallel
Sets CH1 and CH2 to parallel operation mode.
I2+I3 Parallel
Sets CH2 and CH3 to parallel operation mode.
ALL Parallel
Sets CH1, CH2, and CH3 to parallel operation mode.
User Settings
Output Recall
Sets the output state after the Series 2230G is turned on.
Save Last
Sets the voltage, current, and maximum voltage values after the Series
2230G is turned on.
Key Beep
Turns the key beeper on or off.
Knob Lock
Locks the navigation wheel (knob).
Communication Port
GPIB: Sets the communication interface to GPIB. You can set the GPIB
address to a number from 0 to 31 by pressing the up and down arrows.
RS-232: Sets the communication interface to RS-232. You can select the
baud rate (4800, 9600, 19200, 38400, 57600, or 115200) and parity (None,
Even parity, or Odd parity).
USB (default): Sets the communication interface to USB.
System Info
Error Log
Lists all errors that occurred.
Main:0.01/Aux:1.06
Displays the firmware version.
0123456789AF
Displays the serial number.
Basic settings
The following topics describe how to set up and use the Series 2230G.
Set the voltage output or voltage limit for a specific channel
You can set the voltage limit from 0 V to the maximum voltage specified for your Series 2230G.
To set the voltage limit:
1. Select the channel by pressing the
CH1
,
CH2
, or
CH3
key.
2. Press the
V-Set
key.
3. Use the numeric keys, the navigation wheel, or the up, down, right, and left arrow keys to set the
voltage limit.
4. Press the
Enter
key to confirm the setting.