Measurement Queries
:FETCh?
What it does
This command will simply return the latest available reading from an instrument.
Limitations
If the instrument does not have a reading available (indicated by dashes in the display),
sending this command will cause a –230, “Data corrupt or stale” error. This query will not cause
the box to trigger a reading, nor will it “wait” for a result if a reading is in progress. It is possible
to get the same reading over and over using this query. It will continue to give the same result
until one of two things has happened:
•
A new reading has been triggered
•
The old reading has been invalidated by changing ranges, or by changing function.
Where appropriate
Since this query does not trigger a reading, and can give duplicate results, there are not many
cases where this command should be used. The “:DATA:FRESh?” query (
a better choice. If this query is used, the following conditions should be met:
•
A reading has been triggered, either by free running (:INIT:CONT ON and
:TRIG:SOUR IMM), by some event such as a bus trigger (*TRG), or by an external
trigger (:TRIG:SOUR EXT).
•
It is confirmed that the reading is completed, either by the setting of the RAV bit in the
status model, or by allowing sufficient time to pass for the reading to complete.
:READ?
What it does
This command performs three actions. It will reset the trigger model to the idle layer
(equivalent to the :ABORt command), take the trigger model out of idle (equivalent to the :INIT
command), and return a reading (equivalent to a “FETCh?” query). This command will always
return a new reading, since aborting the trigger model will invalidate any old readings and
trigger a new one. This query will “wait” for a new reading to become available before the
instrument sends a result back.
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...