Nanovoltmeter features
The Model 2182 is a 7
1
⁄
2
-digit high-performance digital nanovoltmeter. It has two input
channels to measure voltage and temperature. The measurement capabilities of the Model 2182
are explained in
).
Features of the Model 2182 Nanovoltmeter include:
•
Ratio
— Provides comparison readings between two voltage inputs. Ratio performs
V1/V2.
•
Delta
— Provides average difference of Channel 1 inputs. Delta performs
(V1t1–V1t2)/2.
•
Enhanced Delta, Pulse Delta, and Differential Conductance
— The following tests
can be performed when using a Model 2182/2182A with a Model 6220 or 6221 Current
Source:
—
Delta
- Uses a square wave output and a 3-point measurement algorithm to cancel
the effects of thermal EMFs.
—
Pulse Delta (6221 and 2182A only)
- Provides a pulse output and a 3-point (or
2-point) measurement algorithm for testing of temperature sensitive Device Under
Test (DUT).
—
Differential Conductance
- Uses a differential current output and a 3-point moving
average algorithm to perform differential measurements.
•
mX+b and Percent
— These calculations provide mathematical manipulation of
readings.
•
Relative
— Null offsets or establish baseline values.
•
Buffer
— Store up to 1024 readings in the internal buffer.
•
Limits
— Set high and low reading limits to test devices.
•
Internal Scanning
— Scan the two input channels of the Model 2182.
•
External Scanning
— Scan the channels or matrix points of Keithley Model 7001/7002
switching cards.
•
Setup Storage
— Two instrument setups (user and factory defaults) can be saved and
recalled.
•
Analog Output
— With analog output gain set to one, a full range input will result in a
1V analog output.
•
Remote Interface
— The Model 2182 can be controlled using the IEEE-488 interface
(GPIB) or the RS-232 interface.
•
GPIB Programming Language
— When using the GPIB, the instrument can be
programmed using the SCPI or Model 182 (DDCs) programming language.
•
Closed-cover Calibration
— The Model 2182 can be calibrated from either the front
panel or the GPIB.
1-6
Getting Started
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...