Reference:
A.
ABORt — With continuous initiation disabled, the 2182 goes into the idle state. With
continuous initiation enabled, operation continues at the top of the trigger model.
B.
INITiate — Whenever the instrument is operating within the trigger model, sending this
command causes an error and will be ignored.
C.
INITiate:CONTinuous — With continuous initiation enabled, you cannot use the
READ? command or set sample count (SAMPle:COUNt) greater than one.
D.
FETch? — See
for details on this Signal Oriented Measurement Command.
E.
READ? — Use this one query command to perform the tasks of the three commands.
See
for details on this Signal Oriented Measurement Command.
F.
TRIGger:SOURce — With the timer control source selected, use the TRIGger:TIMer
command to set the interval.
G.
DELay:AUTO — Auto delay period is 5msec for the 100V range, and 1msec for all other
voltage ranges. Disabling auto delay sets the delay time to 0.
H.
TRIGger:SIGNal — Send this action command to bypass the control source when you
do not wish to wait for the programmed event to occur. The instrument must be waiting
at the control source for the event when this command is sent. Otherwise, an error occurs
and the command is ignored.
I.
SAMPle:COUNt — A sample count >1 specifies how many readings will automatically
be stored in the buffer. However, with continuous initiation enabled, you cannot set the
sample count greater than one.
J.
SENSe:HOLD — See “
”, located in this section, for details on
using Hold.
Programming example
The following program fragment triggers (and stores in the buffer) 10 readings, which are
then displayed on the computer CRT.
CALL SEND(7,“*rst”,status%)
‘Restore *RST defaults.
CALL SEND(7,“:trig:del 0.5”,status%)
‘Set delay for 0.5sec.
CALL SEND(7,“:samp:coun 10”,status%)
‘Set sample count to 10.
CALL SEND(7,“:read?”,status%)
‘Trigger and request readings.
reading$ = SPACE$(80)
CALL ENTER(reading$, length%, 7,status%) ‘Address 2182 to talk.
PRINT reading$
‘Display the 10 readings on
‘the CRT.
Triggering
7-17
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...