Figure 1-3
Connections for temperature accuracy verification
2.
Turn on the power, and allow a 2-1/2 hour warm-up period.
3.
Press SHIFT then TCOUPL, then use the RANGE and left and right arrow keys to select
the following:
•
UNITS: C
•
JUNC: SIM, set to 0°C
•
TYPE: J
•
SENS: TCOUPLE
4.
Select the Model 2182 TEMP 1 function.
2182
NANOVOLTMETER
RANGE
!
CHANNEL 1
SHIFT
LOCAL
POWER
RANGE
SHIFT
CH1
REM
TALK
LSTN
SRQ
STAT
REL
FILT
4W
BUFFER
MATH
REAR
SCAN
TIMER
STEP
CH2
CH3
CH4
CH5
CH6
CH7
CH8
CH9
CH10
HOLD
TRIG
FAST
MED
SLOW
AUTO
ERR
AUTO
HI
LO
CHANNEL 2
HI
LO
120V MAX
12V MAX
CAT
I
350V PEAK ANY
TERMINAL TO CHASSIS
EXIT
ENTER
DIGITS RATE
ON/OFF
VALUE
TRIG
EX TRIG
STORE RECALL
DCV1
V
1
/V
2
MX+B
%
V1-V2
DCV2
ACAL
FILT
REL
TEMP
1
TEMP
2
TYPE
AOUT
TCOUPL
DELAY
HOLD
RS232
GPIB
STEP
SCAN
CAL
TEST
OUTPUT
SAVE
RESTR
CONFIG
HALT
BUFFER
SETUP
LIMITS
2107 Cable
Model 2182
Thermocouple
Wires
Connect Wires
in Ice Bath
Thermocouple
Calibrator
L
SYNC
Performance Verification
1-9
Summary of Contents for 2182
Page 10: ......
Page 14: ...1 Performance Verification Performance Verification...
Page 25: ...1 12 Performance Verification...
Page 26: ...2 Calibration Calibration...
Page 43: ...2 18 Calibration...
Page 44: ...3 Routine Maintenance RoutineMain tenance...
Page 47: ...3 4 Routine Maintenance...
Page 48: ...4 Troubleshooting Troubleshoot ing...
Page 64: ...5 Disassembly Disassembly...
Page 73: ...5 10 Disassembly...
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ...6 Replaceable Parts Replaceable Parts...
Page 89: ......
Page 90: ......
Page 91: ......
Page 92: ...A Specifications Specifica tions...
Page 96: ...B CalibrationReference Calibration Reference...
Page 109: ...B 14 Calibration Reference...
Page 110: ...C CalibrationProgram Calibration Program...
Page 115: ...C 6 Calibration Program...
Page 119: ......