Section 3: Basic DMM operation
Model 2110 5½ Digit Multimeter Reference Manual
3-48
2110-901-01 Rev. C/August 2013
Remote operation
Commands that use the
<n>
parameter can use the
MINimum
and
MAXimum
name parameters.
MIN
selects the minimum value and
MAX
selects the maximum value.
Use the following commands to configure and enable the digital filter:
AVERage:TCONtrol <name>
Select filter type
<name>
= MOVing
Moving average filter
= REPeat
Repeating average filter
AVERage:COUNt <n>
Set number of readings to average
<n>
= 2 to 100
Number of readings to average
AVERge:STATe <b>
Enable or disable digital filter
<b>
= 0 or OFF
Disable digital filter
= 1 or ON
Enable digital filter
Math operations
The instrument can perform the following math operations:
•
PERCENT
•
AVERAGE
•
NULL
•
LIMITS
•
mX+b
•
dB
•
dBm
You can apply these math operations to the measurement functions as shown in the following table.
The MATH indicator for the display turns on when a math operation is enabled.
Measurement Function
Math
function
DCV
DCV
RATIO
ACV
DCI, ACI,
Ω
2/
Ω
4, FREQ,
PERIOD,
, TEMP (RTD),
TCOUPL
DIODE
/ CONT
PERCENT
•
•
•
AVERAGE
•
•
•
•
NULL
•
•
•
LIMITS
•
•
•
•
mX+b
•
•
•
•
dB
•
•
dBm
•
•