![Keithley 2000 Calibration Manual Download Page 64](http://html.mh-extra.com/html/keithley/2000/2000_calibration-manual_661840064.webp)
Generating an SRQ on calibration complete
An SRQ (service request) can be used to detect operation complete instead of repeatedly poll-
ing the Model 2000. To use this method, send both *ESE 1 and *SRE 32 to the instrument, then
include the *OPC command at the end of each calibration command line, as covered above. Re-
fer to your controller's documentation for information on detecting and servicing SRQs.
3-20
Calibration Command Reference
Summary of Contents for 2000
Page 1: ...Model 2000Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Page 10: ...1 Performance Verification ...
Page 23: ...2 Calibration ...
Page 45: ...3 Calibration CommandReference ...
Page 65: ...A Specifications ...
Page 76: ...A 12 Specifications ...
Page 77: ...B Error Messages ...
Page 80: ...C Calibration Program ...