Back to Table of Contents 

 

Figure 15: High flow cell fitted to IRmadilloNA probe 

Table 1 

– Specifications and safety limits (high flow cell) 

Parameter 

High Flow Cell 

Pressure 

Refer to high flow cell marking 

Analyte Temperature 

Refer to high flow cell marking 

Body Material 

Corrosion-resistant metal 

O-ring Seal 

FFKM (Kalrez) 

Port Size 

3/4” NPT 

Cell Internal Volume 

~27 ml 

Intended Use 

Only for use with ASM0627-09-N-Cx-O-Gx-D3x 

Fluid Group 

Minimum Inspection Frequency 

Annual 

 

4.2.2.  Low flow cell - ASM1222  

The low flow cell that can be secured to the end of the probe for continuous analysis of your 
process. The flow cell is 

equipped with two 1/16” NPT fittings to attach pipelines too, or it can 

be  assembled 

with  two ¼”  Swagelok nuts  attached to a  short  section  of  pipe  on  opposing 

sides. It is designed to force liquid past the ATR and is primarily used for lab work where flow 
rate and volume are low. 

It is fixed in place using a tri clamp that locates over the rear of the flow cell and two semi-
circular  collars 

‘C-pieces’  that  fit  around  the  probe.  Ensure  that  the  C-pieces  make  good 

contact with the probe and that the pins are seated inside their corresponding holes. Slide the 

Summary of Contents for IRmadillo DOC1049R

Page 1: ...DOC1049R 22 February 2024 12 07 DOC1049R IRmadilloNA User Manual Hardware Model Number ASM0627 09 N Cx O Gx D3x www keit co uk support keit co uk...

Page 2: ...nd 7 2 Maintenance 8 2 1 Spectrum health 8 2 1 1 Use at high temperature 8 2 1 2 Emitter ageing 9 2 2 Remote health check 9 2 3 Cleaning the fibre optic data cable connectors 9 2 3 1 Cleaning the Sour...

Page 3: ...e used The output from this purge gas generator should have a minimum quality according to ISO8573 1 2010 of Class 1 2 1 dust water oil or better Keit can recommend suitable generators that achieve th...

Page 4: ...If the purge tubing is disconnected the blanking plugs must be fitted to the inlet connections on the probe and body to maintain the IP rating and exclude dust The IRmadilloNA requires a stabilisation...

Page 5: ...reduced diameter section of the pipe The minimum flow velocity should be 1 5 m s 5 ft s 4 Install the IRmadilloNA so that the probe tip is fully in the flow We recommend putting the probe tip as close...

Page 6: ...d The following are suitable Water Acetone Alcohols Surfactant solutions Alkanes hexane cyclohexane etc Ethers Once exposed to a suitable solvent wipe the ATR with a clean cloth The metal tube of the...

Page 7: ...of interest within the spectrum is between 800 and 1800 cm 1 This section of the absorbance plot should maintain a relatively flat line with no consistently repeating peaks Cleaning Mode can be used t...

Page 8: ...other background must be taken before any spectra can be measured Figure 6 Negative peaks indicating the background was taken when the probe was unclean or not fully purged 1 4 Taking a background FTI...

Page 9: ...ment 2 1 1 Use at high temperature Figure 7 Probe is heated causing the spectrum to tilt In Figure 7 the spectrum is tilted with positive absorbance at lower wavenumbers and negative absorbance at hig...

Page 10: ...perform an annual examination of your system s health data remotely to determine if any servicing needs to be carried out This includes review of internal diagnostics for warnings and errors component...

Page 11: ...rease the probability of fracture Prot gez la fibre des contraintes et des courbures serr es rayon de courbure 30 mm interdit Les courbures serr es provoquent une contrainte la surface du verre et aug...

Page 12: ...en The pen fits over each of the two male fibre ferrules and is aligned by the transparent cover Push and click once on each ferrule Figure 12 Using the fibre cleaning pen to the Souriau plug Once bot...

Page 13: ...Page 12 DOC1049R 22 February 2024 12 07 www keit co uk Back to Table of Contents Figure 13 Reconnecting the Souriau connectors Locknut Align keyways...

Page 14: ...ions can cause damage to the instrument If in doubt please consult with Keit Le stockage du spectrom tre IRmadilloNA non aliment et non purg dans des conditions inad quates peut endommager l instrumen...

Page 15: ...t force we recommend to tighten the thumbscrew until resistance is felt 1 turn For volatile substances it is recommended to keep the stopper cap in place during use Figure 14 Sample cell fitted to IRm...

Page 16: ...s is an alternative to installing the IRmadilloNA into a main pipeline It is fixed in place using a tri clamp that locates over the rear of the flow cell and two semi circular collars that fit around...

Page 17: ...ency Annual 4 2 2 Low flow cell ASM1222 The low flow cell that can be secured to the end of the probe for continuous analysis of your process The flow cell is equipped with two 1 16 NPT fittings to at...

Page 18: ...ll have not perished They should show no signs of swelling tearing or degradation Also inspects the clamp and C pieces for any signs of degradation cracks or damage If the clamp and or the C pieces sh...

Page 19: ...r as a bench top instrument for offline sampling Angled at 35 the probe can be positioned into beakers or other vessels for ease of use It can also be configured for two heights to give added flexibil...

Page 20: ...meter over such that the mounting holes are pointed upward the instrument should be disconnected from power coms and purge to avoid damage to the cables or twisting Position the inclined stand over th...

Reviews: