3: Custom Calibration 

Warning! Custom Calibration is for advanced users only

This mode can provide the most accurate readings across a wider range than the other calibration 
modes. 

Only proceed if you are comfortable with spreadsheets and complex curves. If you are working with 
solutions with gravities higher than 1.080, then custom calibration is advised. 

For advanced calibration, you will need to take readings at various gravity points roughly 10 gravity 
points apart, and make a note of the angle of the Pill at each point. We recommend readings from 
1.080 down to 1.000, 10 points apart (so, 1.080, 1.070, 1.060 etc). Note that it does not matter if the 
gravity readings are not exactly 10 points apart, as long as the gravity reading is correct. Use of a 
hydrometer that is known to be accurate is quite important. 

To perform Custom Calibration, it is easiest to start with the strongest solution and dilute to 
progressively weaken the solution. It is important to use a calibrated hydrometer to ensure accurate 
gravity readings. 

1.

 

Place the Pill in the first solution of known gravity. Wait for the angle to stabilise, then make 
a note of the angle and the gravity. 

2.

 

Repeat this step for solutions of different gravity, making a note of the angle and the gravity 
at each step. It is easiest to dilute the starting solution in steps of approximately 10 points. 
Finish with a reading in a solution with a gravity of 1.000 (IE water). 

Once you have sufficient readings, the angle and the gravity need to be entered on an Excel 
spreadsheet (or similar such as Google Sheets) as a graph wit

h gravity on the ‘y’ axis and angle on 

the ‘x’ axis. The more readings you take, the more accurate this calibration will be. We recommend a 

minimum of 6 samples, with 10 or more providing a greater degree of accuracy.  

When you have created the graph, create a polynomial trend line onto the graph with the degree of 
the polynomial being 

at least

 one less than the number of data points. If the trend line creates an 

irregular curve (such as not passing through the data point at 1.000 correctly or not following a 
polynomial shape) reduce the number of data points (ensuring that there are still more data points 
than the degree of the polynomial) until the curve matches. 

With 10 data points, a polynomial degree between 5 and 8 will typically give the best results. 

Summary of Contents for RAPT Pill

Page 1: ...APT Pill Hydrometer Thermometer Quick Start Guide Page 1 of 18 www KegLand com au Last Updated 6 05 2022 16 49 PM RAPT PILL Hydrometer Thermometer KL20596 Quick Start Guide KegLand Distribution PTY LT...

Page 2: ...vanced Calibration 11 3 Custom Calibration 12 Updating Firmware 14 Trouble shooting Registration and Telemetry 15 Green light is not illuminating when RAPT Pill plugged in to USB C 15 Red battery ligh...

Page 3: ...KegLand approved cleaners such as StellarOxy and StellarClean and sanitisers such as StellarSan or Phosphoric Acid Battery Installation The battery will come pre installed in the RAPT Pill with a non...

Page 4: ...ure your battery has good contact with the terminals After installing the battery rotate it slightly within the battery holder to ensure it has good contact Important Once you have installed the Li io...

Page 5: ...Fully Charged ESP LED GREEN Slow Green Flashes flashes when the captive portal is enabled and the Pill is searching for a Wi Fi network This will happen when the Pill is plugged in to USB C the first...

Page 6: ...l Captive Portal If you are not prompted to sign into the network of your RAPT Pill by your smart phone or tablet then open your Wi Fi settings and select the RAPT Pill XXXX Wi Fi network again to ope...

Page 7: ...n be used to access the Pill AP in a browser 2 Ensure you are logged in to the RAPT Portal https app rapt io on a second device that is connected to the same home Wi Fi network as the Pill you are reg...

Page 8: ...ortant to insert the circuit board in the correct orientation to ensure the correct weight distribution for accurate readings The yellow end of the circuit board should be inserted into the yellow hou...

Page 9: ...on such as if you are fermenting in a fermentation chamber place the Pill in a jug in the fermentation chamber to confirm that telemetry is received and your RAPT Pill has good signal strength 2 Once...

Page 10: ...ngs and select the RAPT Pill XXXX Wi Fi network again to open the Pill Captive Portal 3 Once the RAPT Pill Access Portal is open on your phone or tablet navigate to the Calibration tab 4 Select the de...

Page 11: ...tion is also quite quick and easy and can provide a more accurate result than Standard Calibration It is the same procedure as standard calibration but in addition to water at 1 000 have a solution of...

Page 12: ...e gravity readings 1 Place the Pill in the first solution of known gravity Wait for the angle to stabilise then make a note of the angle and the gravity 2 Repeat this step for solutions of different g...

Page 13: ...o this y 0 0000023109x5 0 0005741632x4 0 0568746764x3 2 8119846525x2 71 1087310405x 265 6578215178 This is a sample only DO NOT USE THIS EQUATION AS IT WILL NOT BE ACCURATE Substitute x in this equati...

Page 14: ...ur phone or tablet or through the Pill IP address 1 Plug your Pill in to USB power with a USB C cable 2 Connect to the Pill Access Portal with your phone or tablet or through the Pill IP address 3 Nav...

Page 15: ...e Pill stop flashing shortly after it is unplugged This is normal behaviour There is an approximately 10 minute timeout after the Pill is unplugged There are no lights on the Pill when unplugged This...

Page 16: ...o 30dBm Maximum signal 50dBm Excellent signal 67dBm Good reliable signal 70dBm Ok not a strong signal 80dBM Poor unreliable signal 90dBM Unusable Re Register Your RAPT Pill If your RAPT Pill is still...

Page 17: ...perating level the Pill will enter a deep sleep mode in order to protect the battery There may have been an interruption to the network or internet The battery may be slightly loose in the terminals c...

Page 18: ...support please contact your nearest distributor The RAPT Pill Thermometer comes with a 12 month warranty when sold in Australia To lodge a warranty claim in Australia please forward as many visual pie...

Reviews: