3.2.8 Printing data
The measurement results can be printed by pressing the print
button on the hand unit.
This will automatically include the date and time (if set).
A space is included for manually recording the name of
the patient.
The last four individual readings are printed as whole
numbers ‘XX’.
The average IOP is calculated and printed to
one decimal place ‘XX.X’.
Always verify that the data on the printout and data on
the TonoCare Export application matches the readings
on the hand unit.
The printer on the docking station contains a sharp
serrated blade to help cut the paper. Exercise caution to
avoid contact with this blade whenever replacing printer paper
rolls or tearing printouts from the device.
Example:
The TonoCare docking station can also be connected to a USB
port on a PC. It will appear as a serial port in the system.
To receive the measurement data, the port needs to be
configured as follows:
- Baud rate: 115200
- Data bits: 8
- Stop bits: 1
- Parity: none
- Flow control: none
Pressing the print button will send the
results to both the printer and
the USB port.