7. Cleaning
• Only manual non-immersion cleaning as described should be
used for this instrument
• Do not autoclave or immerse in cleaning fluids
Always disconnect power supply from source
before cleaning
a Wipe the external surface with a clean absorbent, non-
shedding cloth dampened with a water / detergent solution
(2% detergent by volume) or water / isopropyl alcohol
solution (70% IPA by volume). Avoid optical surfaces
b Ensure that excess solution does not enter the instrument. Use
caution to ensure cloth is not saturated with solution
c Surfaces must be carefully hand-dried using a clean non-
shedding cloth
d Safely dispose of used cleaning materials