SLIT LAMP

INSTRUCTIONS FOR USE

H Series

A world without vision loss

Summary of Contents for H Series

Page 1: ...SLIT LAMP INSTRUCTIONS FOR USE H Series A world without vision loss...

Page 2: ......

Page 3: ...R USE 11 6 1 SETTING THE BINOCULARS 11 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP 12 6 3 DESCRIPTION OF FILTERS APERTURES AND MAGNIFICATIONS 13 7 ROUTINE MAINTENANCE 14 7 1 LED SYSTEMS 14 7 2 R...

Page 4: ...ty Humidity limitation Use by date Serial number Catalogue number Medical device Translation The Keeler Slit Lamp is designed and built in conformity with Directive 93 42 EEC Regulation EU 2017 745 an...

Page 5: ...embly and a table top with power supply and accessory drawer The light intensity is controlled by a variable rheostat located on the XYZ translation base There are a number of selectable filters allow...

Page 6: ...the specialized technicians of the manufacturer s Technical Service Centre or by personnel trained and authorised by the manufacturer The manufacturer declines any and all responsibility for loss and...

Page 7: ...ortant when the unit has been stored or transported in a cold environment this can cause severe condensation to develop on the optical elements 2 3 CONTRAINDICATION There is no restriction to patient...

Page 8: ...s 9 Power lead chinrest to lamp housing KSL H Series 10 Lamp cover 11 Lamp cover release screws 12 Lever for grey ND blue diffuser and red free filters 13 Slit length slit rotation and aperture contro...

Page 9: ...8 R type tonometer mounting hole 29 Yellow filter knob up out 30 Lock for securing magnification body 31 Magnification change drum 32 Eyepieces adjustable for PD and dioptre correction 33 Breath shiel...

Page 10: ...castors ensure the following before moving it to another location 1 The table is at its lowest position 2 The power cord is removed 3 The Slit Lamp arm and base locking knobs are tightened 4 The runn...

Page 11: ...gned Attach the tower to the base using the hex bolt removed earlier and tighten using the wrench provided 2 Carefully fit the microscope body to the arm ensuring it is pushed to the stop Tighten usin...

Page 12: ...n 9 4 Power Supply page 20 3 Connect the mains power to the Slit Lamp transformer using the power lead provided Only a hospital grade 3 conductor electrical power supply cable must be used For USA and...

Page 13: ...sures exact centering of the prism with the left objective 8 To obtain an image as clear and as free of reflexes as possible the angle between the illumination and the microscope should be about 60 an...

Page 14: ...nus one 1 or minus two 2 dioptres 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP Parts of the slit lamp coming in to contact with the patient should be cleaned in accordance with these instructions...

Page 15: ...mination Tightening the knob will re centre the slit image in the centre of the visual field of the microscope 8 The slit image is made vertical or given a pre set angle by means of the inclination la...

Page 16: ...e outlined below should only be carried out once the main power cable has been disconnected If you have any problems that are not covered by the procedures described below contact Keeler Ltd or your l...

Page 17: ...ables and connectors To access the bulb connections see earlier in this section for guidance 7 5 OPTICS The optics should be wiped clean of any loose dirt or debris with a suitable dust brush then cle...

Page 18: ...g or using this instrument please read carefully and observe what is described here Portable or mobile type radio frequency communication units may have an adverse effect on this instrument resulting...

Page 19: ...1 kV line s to line s 2 kV line s for input output line s 1 kV line s to line s 2 kV line s for input output line s Mains power quality should be that of a typical professional healthcare facility Vol...

Page 20: ...nd people 1 Field strengths from fixed transmitters such as base stations cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predict...

Page 21: ...requency applies These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 9 4 TECHNICAL SPECIFICATIONS O...

Page 22: ...provide at least two means of protection Power Supply Power supply unit Switch mode 100V 240V input 10 multi plug compliant to EN IEC 60601 1 EN IEC 61000 6 2 EN IEC 61000 6 3 Fuse T2 5AH 250V Power s...

Page 23: ...60 hPa 700 hPa TRANSPORT CONDITIONS 50 C 40 C 95 10 1060 hPa 500 hPa Vibration sinusoidal 10 Hz to 500 Hz 0 5g Shock 30 g duration 6 ms Bump 10 g duration 6 ms This instrument does not meet the temper...

Page 24: ...al Keeler Applanation Tonometer 2414 P 2032 Keeler D KAT R Type Digital Keeler Applanation Tonometer 2414 P 2042 11 PACKAGING AND DISPOSAL INFORMATION Disposal of old electrical and electronic equipme...

Page 25: ......

Page 26: ...2021 USA Sales Office Keeler USA 3222 Phoenixville Pike Building 50 Malvern PA 19355 USA Toll Free 1 800 523 5620 Tel 1 610 353 4350 Fax 1 610 353 7814 India Office Keeler India Halma India Pvt Ltd P...