Disposable Probe Adapter
Part Number 2508-P-8026
Instructions for use
The system automatically initiates a purging
cycle of 90 seconds. During the purge cycle an
animated Wait Symbol is displayed alongside
the Probe Symbol.
Three short beeps signal that the purge cycle is
complete. The equipment is now ready for use
as indicated by the timer and the Ready Symbol.
Freezing cycles can now be controlled by the Cryomatic
footswitch as normal (see Cryomatic user manual).
Warnings and Precautions
• Only to be used in conjunction with Keeler
Cryomatic (2509-P-1000) and Disposable Cryo
Probe (2508-P-7022)
• Check probe before use – do not use or repair
if damaged or deformed
• Disconnect probe from adapter only when
system is depressurised
• Maximum pressure 5860kPa (850psi)
• Consult Cryomatic instruction manual if in doubt
• Unauthorised interference invalidates warranty
• This device should be used by trained staff
experienced in the use of ophthalmic equipment
• USA federal law restricts this device to sale or
on the order of a physician
1
Connect the Probe Adapter to the Cryomatic console
by aligning the coupling on the console spigots
and pushing to a positive click (It is impossible to
connect the coupling the wrong way round – correct
orientation is with the release button uppermost).
Connect the disposable probe by inserting the probe
plug into the adapter and pushing the probe firmly
and squarely against the collar(1).
The safety latch(2) will operate and the locking
ring(3) will spring towards the probe to confirm that
the probe has been successfully connected.
Operation
2
Before use ensure that the rear safety
clip(4) is in the fully down position
- under no circumstances should it
be used if the safety clip is not in
position (refer to image on page 2).
1
2
4
3
Sterilized using Ethylene Oxide
Manufacturer
Caution
The CE mark on this product indicates
it has been tested to and conforms
with the provisions noted within the
93/42/EEC Medical Device Directive
Do Not Reuse
Use By
Batch Code/Lot Number/Batch Number
Consult Instructions For Use
This Symbol on the Product or on its
Packaging and instructions indicates
that it was put on the market place
after August 2005 and that this
product shall not be treated as
Household Waste. EU only.
Symbols
3
0088
Continued overleaf