SHENZHEN KEDIDA ELECTRONICS CO., LTD
1
http://www.kedida.com.cn/
Salt Meter Instruction Manual
CT-3081
Thanks for choosing our products. Please read the manual carefully before using the meter. And keep this manual in
a safe place for future reference. Always rinse the electrode in distilled water for 10 mins. and wipe carefully with a
tissue before use.
FEATURES
* 0 to 9999 mg/L
* LCD indicates both mg/L & Temp.
* Auto calibration for 6640 mg/L
(
12.88 ms/cm
)
* Data hold function for freezing the desired value
* Water proof
* Auto power off without operation in 5 mins
* Easy to change the salt meter electrode
*Microcomputer circuit, intelligent function, high
accuracy
* Temperature measurement,
℃
or
℉
* Compact size, light weight
* Build in temperature sensor, ATC (auto temperature
compensation)
* Power supply by 1.5 V battery (LR44) x 4 pcs
* Pen type digital salt meter, all in one, salt electrode is
included, easy for general purpose application
*Available for wide applications, such as aquarium,
beverage, fish hatcheries, food processing, laboratory,
quality control, school & colleges, swimming pools,
water conditions
SPECIFICATIONS
Display
LCD, size : 20 mm x 27 mm. Consumption
Measurement
Range
salt
0 to 9999 mg/L
Temp.
0 to 50
℃
( 32 to 122
℉
)
Resolution
salt
1 mg/L
Temp.
0.1
℃
, 1
℉
Accuracy
salt
± 5 mg/L+1%FS
* After calibration
Temp.
± 0.5
℃
, ± 1
℉
salt Calibration
6640 mg/L ,
(
12.88 ms/cm
)
1 point calibration
Operating Humidity
Less than 80% RH
Dimension
188 x 38 mm (electrode included).
Weight
82 g (electrode included).