- 16 -
Chamber Lid Seal Ring
●
The dirt on the seal ring may not allow you to
reach sufficient vacuum level or vacuum may
leak during pressure application. Use a clean
wet cloth to wipe the dirt on the seal ring
before casting operation.
Caution & Warning
■
When the following messages are displayed, address according to the message. Contact our service
team if there is no improvement. The following errors may sometimes be displayed in cases when there
is extreme interference. Please turn off the unit and on again to see if the error code persists.
Note If the errors are not displayed after the unit is turn off and on again, there may be deletion or modification of casting
)
programs and defaults sets.
The followings are considered to be possible source of interference.
Lighting strikes,
Arc casting machine in the vicinity,
Radio transmitter in the vicinity,
・
・
・
Induction motor in the vicinity.
・
Cautions
●
When the following cautions are displayed, you can skip cautions to use the unit. However we recommend
※
you to maintain and check the unit when errors/cautions are displayed to prevent further malfunctions.
1.
Time over of induction
High frequency current carrying time reaches the limited time.
The limited time is 5 minutes. Press any button to return to normal display.
2.
Abnormal cooling fan filter
The cooling fan does not work properly. Clean the filter at the bottom of the unit.
When there is no improvement, replacement of the fan may be required.
Press any button to return to normal display. However we recommend you to maintain the unit promptly.
3.
Water Overheating
When using the optional chiller, chiller water temperature is overheating.
Wait for 10 minutes to use the unit. When it is often displayed than before, clean the chiller filter.
If room temperature is high, there is a high possibility of this caution.
※
Seal Ring