3 Replacing and checking cleaning and germ reduction agents | 3.3 Replenishing OXYGENAL 6 when using the wa-
ter bottle
52 / 56
3.3 Replenishing OXYGENAL 6 when using the water
bottle
Dosing attachment: (
Mat. no. 1.002.0287
)
Filling the water bottle with the dosing device
NOTE
Before you use the OXYGENAL 6, check the expiry date of the product.
4
Wipe up any spilled OXYGENAL immediately in order to prevent discoloura-
tion of the surfaces.
4
Remove the protective hose from the tube of the KaVo OXYGENAL 6 dosing
device.
4
Guide the tube through the orifice of the coarse sieve of the KaVo OXY-
GENAL 6 bottle.
4
Screw the KaVo OXYGENAL 6 dosing device tightly to the KaVo OXYGENAL
6 bottle.
4
Rotate the orifice in the lid to a filling level of 1.5 litres.
ð
Lid snaps-in in the filling position.
4
Place the KaVo OXYGENAL 6 bottle on a level surface.
4
Compress the KaVo OXYGENAL 6 bottle until the germ reduction liquid rises
above the red marker.
ð
In response to over-filling, the excess quantity above the tick mark is
suctioned off automatically such that exact dosing is ensured.
4
Place the KaVo OXYGENAL 6 bottle in the water bottle. Make sure that the
spout of the KaVo OXYGENAL 6 dosing device (nose) is situated right above
the opening of the water bottle.