Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.7 Reprocessing the three-function and multifunctional
handpiece
21 / 56
2.7.1 Preparations at the site of use
WARNING
Hazard from contaminated products.
Contaminated products are associated with an infection hazard.
4
Take suitable personal protective measures.
CAUTION
Danger of injury if the valve body is hot (heating elements, high-pres-
sure lamp).
Risk of burn injury.
4
Switch off the unit at the main switch.
4
Allow the handpiece to cool down after long use.
4
Remove gross soiling immediately.
4
Reprocess the medical device right after treatment.
4
Leave the gripping sleeve and cannula on the instrument, and activate the
air and water for 60 seconds.
4
The medical device must be dry when transported to reprocessing.
4
To minimise the risk of infection during reprocessing, always wear personal
protective equipment.
4
Do not place in solutions or similar substances.
2.7.2 Disassembly
WARNING
Incomplete reprocessing.
Infection hazard.
4
To ensure complete reprocessing of all parts, the medical device needs to be
disassembled before reprocessing.
4
Hold the three-function or multifunctional handpiece by the gripping sleeve
②
and take off the cannula
④
with a slight twisting motion.
4
Pull the gripping sleeve
②
off the valve body
①
.
4
Detach the key part
③
from the gripping sleeve
②
.