Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.7 Reprocessing the three-function and multifunctional
handpiece
20 / 56
Cleaning and disinfection of the suction hose connector
4
The manual external cleaning and disinfection of the connector involves
manual cleaning of the exterior and wipe disinfection.
4
The manual internal cleaning and disinfection is part of the cleaning and dis-
infection of the suction system using DEKASEPTOL Gel. Keep the connectors
on the suction hoses for this purpose.
or
4
Subject the disassembled parts to automated cleaning and disinfection.
4
Parts should be standing upright when placed in the washer disinfector.
NOTE
Reprocess the suction cannulas according to the instructions of the manufac-
turer.
2.6.4 Reprocessing of the sieve inserts and suction
hoses
NOTE
Check the sieve inserts daily and after each surgery. The sieve inserts (
Mat.
no. 0.763.3423
) must be replaced at the latest when the suction perform-
ance decreases. Cleaning is not recommended for hygienic reasons.
4
Hold the napkin under the sieve seat, and pull the sieve insert out of the
housing.
4
If needed, insert the new sieve insert (
Mat. no. 0.763.3423
) into the seat.
4
Subject the suction hose is to wipe disinfection.
2.7 Reprocessing the three-function and multifunctional
handpiece
NOTE
Wet the O-rings of the gripping sleeves after sterilisation using silicone grease
(
Mat. no. 1.000.6403
). After replacing the cannula, the air duct must be
blown out before the start of treatment to ensure that the air exiting during
treatment is dry.