2 Reprocessing steps in accordance with ISO 17664 | 2.4 Preparation of the dentist element
14 / 56
4
Also refer to: Chapter on Automatic cleaning programmes
4
Pull the latches outward evenly and carefully remove the handpiece holder.
4
Clean the handpiece holder manually.
4
Subject the handpiece holder to wipe disinfection.
2.4.3 Reprocessing of the swing arm S table
Cleaning and disinfection
NOTE
The swing arm S-table (
Mat. no. 1.007.4755
) can be removed to facilitate
cleaning.
NOTE
No cleaning programmes can be run when the swing arm is removed.
4
Locking the screen.
4
Take the handpieces off the hoses.
4
Take the roller out of the swing arm and then place the hose in the hygiene
centre.