Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.4 Preparation of the dentist element
13 / 56
2.4 Preparation of the dentist element
2.4.1 Reprocessing of the handle and holder pad / non-
slip mat
Cleaning and disinfection
4
Press the push-button and pull off the handle (S-table and T-table only).
4
Clean the handle manually.
4
Disinfect the white handle by wiping and the grey handle by wiping or
thermal disinfection (minor discolouration may occur).
4
Sterilise the holder pad and non-slip mat in a washer disinfector.
4
Parts should be standing upright when placed in the washer disinfector.
Sterilisation
4
Sterilise the holder pad/ non-slip mat.
4
The grey handle (
Mat. no. 1.002.4489
) can be sterilised (see symbol on
the handle).
2.4.2 Reprocessing of handpiece holder
Cleaning and disinfection
4
Place the instrument hoses in the designated places of the hygiene centre
during the cleaning and disinfection of the handpiece holder.