Instructions for use SURGmatic S201 XL Pro - 1.013.7541 SURGmatic S201 L Pro - 1.013.7540
7 Processing steps in accordance with EN ISO 17664 | 7.3 Manual processing
18 / 28
7.3 Manual processing
WARNING
Sharp tool in the medical device.
Injury hazard from sharp and/or pointed tool.
▶
Remove the tool.
NOTICE
Never process this medical device in an ultrasonic device.
Functional damage and property damage.
▶
Clean it in a washer disinfector only.
Note
The spray clip for internal cooling may be additionally cleaned in the ultra-
sonic cleaner.
7.3.1 Manual internal and external cleaning and internal
and external disinfection
Not applicable.
7.3.2 Manual drying
▶
Clean the outside and inside with compressed air until no drops of water are
visible.
▶
Immediately after drying, lubricate the KaVo medical device with care
agents from the KaVo care system.
See also:
2
7.5 Care products and systems - Servicing, Page 20
7.4 Automated processing
WARNING
Incomplete disinfection.
Infection hazard.
▶
Only use disinfection procedures that are verified to be bactericidal, fungici-
dal and virucidal.
▶
If the disinfectants used do not meet these requirements, the process must
be concluded by disinfection of the unit(s) without packaging using a steril-
izer.