Operating Instruction STERIclave B 2200 / 2200 P
6 Operation
|
6.4 Selecting and starting a program
Display
Program
Total cycle test
(Including drying)
Parameter
Load
(test according to DIN EN
13060:2004)
Type
(DIN EN 13060:2004)
Test
Bowie & Dick test 16 min
134 ℃
3.5 min
3 pre-vacuum pha‐
ses
For B&D tests
(3M Comply TM,
Kode 1300)
Vacuum test
16 min
With a cold device
below 35℃ from
standby
For a daily seal
test
*Observe country-specific regulations. Deviating parameters may be adjustable in
the SPECIAL 4 program and executive bowl with this program.
Note
The program Special S4 does not correspond to type B sterilisation.
See also: 6.5 Programming and starting Special S4, Page 48
6.4.2 Starting Sterilize 1
▶ Press 1 to select the Sterilize 1 program.
The selected program will appear for 5 seconds on the pressure display.
▶ Press the Start/Stop button to start the selected program.
The sterilisation cycle starts. The program sequence LEDs light up to indicate
which phase of the cycle is currently active.
See also: 6.6 Tracking the program sequence, Page 50
6.4.3 Starting Sterilize 2
▶ Press 2 to select the Sterilize 2 program.
The selected program will appear for 5 seconds on the pressure display.
▶ Press the Start/Stop button to start the selected program.
The sterilisation cycle starts. The program sequence LEDs light up to indicate
which phase of the cycle is currently active.
See also: 6.6 Tracking the program sequence, Page 50
6.4.4 Start fast cycle
▶ Press 3 to select the Fast cycle program.
The selected program will appear for 5 seconds on the pressure display.
+
▶ Press 3 and the Start/Stop button at the same time to start the selected program.
The sterilisation cycle starts. The program sequence LEDs light up to indicate
which phase of the cycle is currently active.
See also: 6.6 Tracking the program sequence, Page 50
Note
The display shows the most important program parameters for the selected pro‐
gram.
46/78