Instructions for use SONICflex seal tips

6 Reprocessing steps in accordance with ISO 17664 | 6.4 Packaging

17 / 20

NOTE

Please comply with the instructions for use of the washer disinfector.

6.4 Packaging

NOTE

The sterile goods package must be large enough to accommodate the product
without stretching the packaging. The quality and use of the packaging of the
items to be sterilised must meet the applicable standards and be appropriate
for the sterilisation process!

4

Seal the medical device separately in a sterile pack.

6.5 Sterilisation

Sterilisation in a steam steriliser (autoclave) in
accordance with EN 13060 / EN ISO 17665-1

CAUTION

Contact corrosion due to moisture.

Damage to product.

4

Immediately remove the product from the steam steriliser after the steril-
isation cycle.

The medical device has a max. temperature resistance of up to 138 

o

C (280.4

o

F).

Sterilisation parameters:

Select a suitable process from the following sterilisation processes (depending
on the available steriliser):

▪ Steriliser with triple pre-vacuum:

– at least 3 minutes at 134 

o

C -1 

o

C / +4 

o

C (273 

o

F -1.6 

o

F / +7.4 

o

F)

▪ Steriliser using the gravity method:

– at least 10 minutes at 134 

o

C -1 

o

C / +4 

o

C (273 

o

F -1.6 

o

F / +7.4 

o

F)

– at least 60 minutes at 121 

o

C -1 

o

C / +4 

o

C (250 

o

F -1.6 

o

F / +7.4 

o

F)

Use according to the manufacturer's Instructions for Use.

6.6 Storage

Reprocessed products must be stored appropriately protected from light in a
dry, dark, cool low-germ room.

NOTE

Comply with the expiry date of the sterilised items.

Summary of Contents for SONICflex

Page 1: ...Instructions for use SONICflex seal tip no 45 1 000 8323 SONICflex seal tip no 45A 1 007 1506 SONICflex seal tips...

Page 2: ...Distribution KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Page 3: ...tartup and shut down 11 4 1 Inserting the SONICflex tips 11 4 2 Removing the SONICflex tip 12 5 Operation 13 5 1 Setting the power on the SONICflex 13 5 2 Notes for application 13 6 Reprocessing steps...

Page 4: ...Service 49 0 7351 56 1000 service instrumente kavokerr com Target group The instructions for use are intended for medical professionals in particular dentists and dental practice personnel The sectio...

Page 5: ...ying documents EAC conformity mark Eurasian Conformity Follow the electronic instructions for use GOST R certification HIBC Code Material number Manufacturer Serial number Transportation and storage c...

Page 6: ...y and material damage The warning notes are designated as shown below HAZARD In cases which if not prevented directly lead to death or severe injury WARNING In cases which if not prevented can lead to...

Page 7: ...Technical condition A damaged device or components could injure patients users and third parties 4 Only operate devices or components if they show no signs of damage on the outside 4 Check that the de...

Page 8: ...ion with the product by the manufacturer 4 Only use accessories that are equipped with standardised interfaces 4 Do not make any modifications to the device unless these have been ap proved by the man...

Page 9: ...gulations the user is required to only use equipment that is working correctly to comply with the specified intended use to protect him or herself the patient and third parties from hazards to prevent...

Page 10: ...ortation and storage conditions CAUTION Startup after refrigerated storage Malfunction 4 Prior to startup strongly refrigerated products must be allowed to warm up to a temperature of 20 C to 25 C 68...

Page 11: ...Inserting the SONICflex tips CAUTION Tip incorrectly inserted in the torque wrench Danger of injury to the user 4 When inserting the tip into the torque wrench make sure that the end of the tip always...

Page 12: ...for use SONICflex seal tips 4 Startup and shut down 4 2 Removing the SONICflex tip 12 20 4 2 Removing the SONICflex tip 4 Place the torque wrench on the SONICflex tip and unscrew it in counter clockwi...

Page 13: ...EVEL 3 brief CAUTION Hazard from damage to the enamel Any extended use of level 3 is associated with the risk of damage to the enamel and premature tip breakage 4 Level 3 should only be used briefly 4...

Page 14: ...stallisation in the tips 4 Reprocess the medical device as soon as possible after treatment 4 The medical device must be dry when transported to reprocessing 4 To minimise the risk of infection during...

Page 15: ...infection bath or chloride containing disinfectants Malfunction and material damage 4 Do not disinfect the device in the disinfection bath or with chloride contain ing disinfectants KaVo recommends th...

Page 16: ...ips long thread 1 009 7875 Service adaptor SONICflex quick tips short thread 1 009 7876 Service adaptor SONICflex tip 4 Use the wrench to screw the SONICflex tip onto the service adaptor in clockwise...

Page 17: ...rrosion due to moisture Damage to product 4 Immediately remove the product from the steam steriliser after the steril isation cycle The medical device has a max temperature resistance of up to 138 o C...

Page 18: ...ers Material summary Material number Torque wrench 1 000 4887 Nozzle pin 0 410 0911 Service adaptor Material summary Material number Service adaptor SONICflex 2003 tips long thread 1 009 7875 Service...

Page 19: ......

Page 20: ...1 000 8325 bd 20211130 03 en...

Reviews: