Instructions for use SONICflex seal tips
6 Reprocessing steps in accordance with ISO 17664 | 6.2 Manual Reprocessing
15 / 20
6.2.2 Manual internal cleaning
This product is not designed for manual internal cleaning. For effective repro-
cessing, the inside of the device must be subjected to automated cleaning in a
cleaning and disinfection unit in accordance with EN ISO 15883-1.
6.2.3 Manual external disinfection
WARNING
Incomplete disinfection.
Infection hazard.
4
Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
4
If the disinfectants used do not meet these requirements, the process must
be concluded by disinfection of the unit(s) without packaging using a steam
steriliser.
CAUTION
Use of the disinfection bath or chloride-containing disinfectants.
Malfunction and material damage.
4
Do not disinfect the device in the disinfection bath or with chloride-contain-
ing disinfectants.
KaVo recommends the following products based on the compatibility of the ma-
terials. The microbiological efficacy must be ensured by the disinfectant manu-
facturer and proven by an expert opinion.
Approved disinfectants:
Consumables required:
▪ Cloths for wiping the medical device.
4
Spray the disinfectant on a cloth, then wipe down the medical device and al-
low the disinfectant to act according to the instructions of the disinfectant
manufacturer.
4
Follow the instructions for use of the disinfectant.
6.2.4 Manual internal disinfection
This product is not designed for manual internal disinfection. For effective re-
processing, the inside of the device must be subjected to automated cleaning in
a cleaning and disinfection unit in accordance with EN ISO 15883-1.