Instructions for use SONICflex rootplaner 

6 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 6.4 Care products and systems - Servicing

17 / 20

4

Use the torque wrench to screw the SONICflex tip 

 to cleaning adapter 

in a clockwise direction.

4

Remove the torque wrench.

4

Attach cleaning adapter 

 to Miele adapter AUF 1 in combination with ADS

1.

KaVo recommends washer disinfectors in accordance with EN ISO 15883-1,
which are operated using alkaline cleaning agents.

The validations were performed using a Miele washer disinfector and the
"VARIO-TD" programme and a mild alkaline cleaner made by Dr. Weigert.

In addition, KaVo recommends the use of a rinsing agent.

4

For programme settings as well as cleansers and disinfectants, please refer
to the Instructions for Use of the washer disinfector.

4

For adaptations, unless otherwise specified, refer to the Instructions for Use
of the washer disinfector.

6.3.3 Automated drying

The drying procedure is usually part of the cleaning programme of the washer
disinfector.

NOTE

Please comply with the instructions for use of the washer disinfector.

4

In order not to affect the KaVo medical device, make sure that the product
is dry on the inside and outside after completion of the cycle.

6.4 Care products and systems - Servicing

Not applicable.

6.5 Packaging

NOTE

The sterile goods package must be large enough to accommodate the product
without stretching the packaging. The quality and use of the packaging of the
items to be sterilised must meet the applicable standards and be appropriate
for the sterilisation process!

4

Seal the medical device separately in a sterile pack.

6.6 Sterilisation

Sterilisation in a steam steriliser (autoclave) in
accordance with EN 13060 / EN ISO 17665-1

CAUTION

Contact corrosion due to moisture.

Damage to product.

4

Immediately remove the product from the steam steriliser after the steril-
isation cycle.

The medical device has a max. temperature resistance of up to 138 

o

C (280.4

o

F).

Summary of Contents for SONICflex rootplaner 24

Page 1: ...A 1 006 1957 SONICflex rootplaner tip no 25 0 571 5631 SONICflex rootplaner tip no 25 A 1 006 1959 SONICflex rootplaner tip no 26 0 571 5641 SONICflex rootplaner tip no 26 A 1 006 1961 SONICflex root...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Page 3: ...tion and storage conditions 10 4 Startup and shut down 11 4 1 Inserting the SONICflex tips 11 4 2 Removing the SONICflex tip 12 5 Operation 13 5 1 Setting the power on the SONICflex 13 5 2 Notes for a...

Page 4: ...om KaVo Technical Service If you have any technical questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service instrumente kavo com Target group The instructions for u...

Page 5: ...e accompanying documents Follow the electronic instructions for use HIBC Code UDI symbol CE mark for medical devices Medical device labelling of medical devices Transportation and storage conditions t...

Page 6: ...y and material damage The warning notes are designated as shown below HAZARD In cases which if not prevented directly lead to death or severe injury WARNING In cases which if not prevented can lead to...

Page 7: ...injury to soft tissue do not continue treatment in the oral cavity with instruments driven by compressed air 4 Place the torque wrench on the tip when the handpiece is in its holder 2 2 Technical cond...

Page 8: ...e accessories that have been approved for combination with the product by the manufacturer 4 Do not make any modifications to the device unless these have been ap proved by the manufacturer of the pro...

Page 9: ...and third parties from danger Contamination by the product must be avoided According to these regulations the medical device is to be used exclusively for the described application while complying wi...

Page 10: ...th long thread Tips without A can be used for all versions of KaVo SONICflex 2003 KaVo SONICflex 2000 Tips with short thread Tips with A can be used for all versions of KaVo SONICflex quick 2008 3 5 T...

Page 11: ...fer to 6 Reprocessing steps in accordance with ISO 17664 1 ISO 17664 2 Page 15 Current packaging law Dispose of and recycle the packaging appropriately in accordance with current packaging law employi...

Page 12: ...by turning it clockwise 4 Use the thin region of the handle for insertion 4 Use the thick region of the handle for tightening or detachment NOTE The tip is properly tightened when the torque wrench s...

Page 13: ...prevention of additional periodontal degradation are primary This especially hold true for the furcations To prevent undesired loss of local substance the tips should be continuously guided back and f...

Page 14: ...s two mirrored contra angle attachments tip no 24 24 A 25 25 A Do not process enamel or filling margins since they could quickly blunt the at tachments The diamond coated airscaler attachments are onl...

Page 15: ...reatment 4 Remove all residual cement composite or blood immediately 4 Wipe disinfect the medical device before transport 4 Remove the tip from the medical device 4 Do not place in solutions or simila...

Page 16: ...al damage 4 Reprocess it in a washer disinfector only CAUTION Never reprocess this medical device in an ultrasonic cleaner Malfunction and material damage 4 Reprocess it in a washer disinfector only 6...

Page 17: ...f the washer disinfector 6 3 3 Automated drying The drying procedure is usually part of the cleaning programme of the washer disinfector NOTE Please comply with the instructions for use of the washer...

Page 18: ...6 o F 7 4 o F Steriliser using the gravity method at least 10 minutes at 134 o C 1 o C 4 o C 273 o F 1 6 o F 7 4 o F at least 30 minutes at 121 o C 1 o C 4 o C 250 o F 1 6 o F 7 4 o F 4 Remove the med...

Page 19: ...om specialised dental dealers Consumables Material summary Material number Torque wrench 1 000 4887 Nozzle needle 0 410 0911 Cleaning adapter Material summary Material number Cleaning adapter for SONI...

Page 20: ...0 589 1722 s 20240306 03 en...

Reviews: