KaVo SONICflex rootplaner 24 Instructions For Use Manual Download Page 11

Instructions for use SONICflex rootplaner 

4 Startup and shut down | 4.1 Inserting the SONICflex tips

11 / 20

4 Startup and shut down

CAUTION

Premature wear and malfunctioning from improper storage during
long periods of nonuse.

Reduced product life.

4

The medical device should be cleaned, serviced and stored in a dry location,
according to instructions, before long periods of nonuse.

 WARNING

Dispose of the product in appropriate manner.

Patients, users or third parties could be infected by contaminated medical
devices.

4

Reprocess the product and accessories before disposal.

 WARNING

Hazard from contaminated products.

Patients, users or third parties could be infected by contaminated medical
devices.

4

Prior to initial startup and after each use, reprocess the product and ac-
cessories.

Also refer to: 

6 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2, Page 15

Current packaging law

Dispose of and recycle the packaging appropriately in accordance with current
packaging law, employing waste management / recycling companies. Comply
with the comprehensive return system. KaVo has had its packaging licensed for
this purpose. Please comply with the regional public waste-disposal system.

4.1 Inserting the SONICflex tips

 WARNING

Tip falls off during the treatment.

Unless the tip has been screwed in properly, it can become detached during the
treatment and lead to injuries.

Aspiration, swallowing of parts and danger of suffocation.

4

Use the original SONICflex torque wrench.

4

Use the original SONICflex tips.

4

Screw in SONICflex tips until you hear a click.

 WARNING

Use of non-approved tips.

Injury to the patient or damage to the medical device.

4

Comply with the instructions for use and the intended use of the tips.

4

Only use tips that do not deviate from the specified data.

Summary of Contents for SONICflex rootplaner 24

Page 1: ...A 1 006 1957 SONICflex rootplaner tip no 25 0 571 5631 SONICflex rootplaner tip no 25 A 1 006 1959 SONICflex rootplaner tip no 26 0 571 5641 SONICflex rootplaner tip no 26 A 1 006 1961 SONICflex root...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Page 3: ...tion and storage conditions 10 4 Startup and shut down 11 4 1 Inserting the SONICflex tips 11 4 2 Removing the SONICflex tip 12 5 Operation 13 5 1 Setting the power on the SONICflex 13 5 2 Notes for a...

Page 4: ...om KaVo Technical Service If you have any technical questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service instrumente kavo com Target group The instructions for u...

Page 5: ...e accompanying documents Follow the electronic instructions for use HIBC Code UDI symbol CE mark for medical devices Medical device labelling of medical devices Transportation and storage conditions t...

Page 6: ...y and material damage The warning notes are designated as shown below HAZARD In cases which if not prevented directly lead to death or severe injury WARNING In cases which if not prevented can lead to...

Page 7: ...injury to soft tissue do not continue treatment in the oral cavity with instruments driven by compressed air 4 Place the torque wrench on the tip when the handpiece is in its holder 2 2 Technical cond...

Page 8: ...e accessories that have been approved for combination with the product by the manufacturer 4 Do not make any modifications to the device unless these have been ap proved by the manufacturer of the pro...

Page 9: ...and third parties from danger Contamination by the product must be avoided According to these regulations the medical device is to be used exclusively for the described application while complying wi...

Page 10: ...th long thread Tips without A can be used for all versions of KaVo SONICflex 2003 KaVo SONICflex 2000 Tips with short thread Tips with A can be used for all versions of KaVo SONICflex quick 2008 3 5 T...

Page 11: ...fer to 6 Reprocessing steps in accordance with ISO 17664 1 ISO 17664 2 Page 15 Current packaging law Dispose of and recycle the packaging appropriately in accordance with current packaging law employi...

Page 12: ...by turning it clockwise 4 Use the thin region of the handle for insertion 4 Use the thick region of the handle for tightening or detachment NOTE The tip is properly tightened when the torque wrench s...

Page 13: ...prevention of additional periodontal degradation are primary This especially hold true for the furcations To prevent undesired loss of local substance the tips should be continuously guided back and f...

Page 14: ...s two mirrored contra angle attachments tip no 24 24 A 25 25 A Do not process enamel or filling margins since they could quickly blunt the at tachments The diamond coated airscaler attachments are onl...

Page 15: ...reatment 4 Remove all residual cement composite or blood immediately 4 Wipe disinfect the medical device before transport 4 Remove the tip from the medical device 4 Do not place in solutions or simila...

Page 16: ...al damage 4 Reprocess it in a washer disinfector only CAUTION Never reprocess this medical device in an ultrasonic cleaner Malfunction and material damage 4 Reprocess it in a washer disinfector only 6...

Page 17: ...f the washer disinfector 6 3 3 Automated drying The drying procedure is usually part of the cleaning programme of the washer disinfector NOTE Please comply with the instructions for use of the washer...

Page 18: ...6 o F 7 4 o F Steriliser using the gravity method at least 10 minutes at 134 o C 1 o C 4 o C 273 o F 1 6 o F 7 4 o F at least 30 minutes at 121 o C 1 o C 4 o C 250 o F 1 6 o F 7 4 o F 4 Remove the med...

Page 19: ...om specialised dental dealers Consumables Material summary Material number Torque wrench 1 000 4887 Nozzle needle 0 410 0911 Cleaning adapter Material summary Material number Cleaning adapter for SONI...

Page 20: ...0 589 1722 s 20240306 03 en...