6 Preparation methods according to ISO 17664 | 6.6 Packaging
▶ Place the product on the cannula, and press the spray button for one second.
6.5.3 Care of the KaVo SPRAYrotor
▶ Remove tip
▶ Place the product on the appropriate coupling of the KaVo SPRAYrotor and cover
it with a Cleanpac bag.
▶ Servicing the product.
See also:
2
Instructions for use KaVo SPRAYrotor
6.5.4 Care with KaVo QUATTROcare
Cleaning and servicing unit with expansion pressure for effective cleaning and care.
KaVo recommends servicing the product once per week.
▶ Remove the tip.
▶ Servicing the product.
See also:
2
Instructions for use KaVo QUATTROcare 2104 / 2104A
6.6 Packaging
Note
The sterilisation bag must be large enough for the handpiece so that the bag is not
stretched.
The quality and use of the sterilisation packaging must satisfy applicable standards
and be suitable for the sterilisation procedure!
▶ Weld each medical device individually in a sterilised item package!
6.7 Sterilisation
Sterilisation in a steam steriliser (autoclave) in accordance with EN
13060 / ISO 17665-1
CAUTION
Premature weary and malfunctions from improper servicing and care.
Reduced product life.
▶ Before each sterilisation cycle, service the medical device with KaVo care prod‐
ucts. However, if over-oiled, the SONICflex will differ from the guidelines.
19 / 26