5 Operation | 5.1 Attaching the SONICflex
5 Operation
5.1 Attaching the SONICflex
CAUTION
Ensure that the SONICflex is tightly seated on the coupling.
If the SONICflex unintentionally comes off the coupling during treatment, it can en‐
danger the patient and user.
▶ Before each treatment, pull on the SONICflex to see if it is securely seated in the
coupling.
CAUTION
Imprecise coupling especially during the afterglow period.
Imprecise coupling can destroy the high-pressure bulb of a MULTIflex (LUX) coupling
or reduce its life.
▶ Make sure that the coupling is precise.
▶ Accurately position the SONICflex on the MULTIflex (LUX)/MULTIflex LED cou‐
pling and press it to the rear until it audibly locks.
5.2 Removing the SONICflex
▶ Hold the MULTIflex (LUX)/ MULTIflex LED coupling tight, and pull the SONICflex
forward while twisting slightly.
5.3 Inserting the SONICflex tip
CAUTION
Hazard from a tip which has been incorrectly inserted in the torque wrench.
This may result in a risk of injury for the user.
▶ When inserting the tip into the torque spanner, make sure that the tip's end al‐
ways faces into the opening of the torque spanner.
Note
In view of the substantial liability risks, we recommend using exclusively original Ka‐
Vo SONICflex tips.
▶ Insert the desired tip with the tip end pointing down into the torque wrench and
screw it into the handpiece by turning it clockwise.
The torque wrenches are for changing the working tips of the SONICflex and protect
against injury. The torque wrench can be screwed in more quickly by holding it at the
rear, thin grip area ①. The large diameter ② helps tighten and remove it.
Note
The tip is properly gripped when the torque wrench slips.