You can now save the suction level.
▶ Turn off the power switch ⑤, and turn it on after two seconds.
The selected suction level is saved.
Deactivate the default suction level
▶ Hold down the "Minus" ⑥ key, and turn the power switch ⑤ on.
The LEDs ① to ④ flash.
▶ Release the "Minus" ⑥ key.
The suction level is now saved.
▶ Turn off the power switch ⑤, and turn it on after two seconds.
The selected suction level is saved.
6.6
Setting the response sensitivity
Prepare the unit
The response sensitivity is set in the "Automatic" position.
I
0
3
nmax
x1000/min
K
con trol
GAS
DIN
DVG
W
SMARTair
▶ Switch the power switch ③ to "On".
▶ Hold down the "Automatic" key ④ (for about three seconds), until the LEDs ①
and ② flash.
The unit is now in "Setting mode".
Instructions for use SMARTair/plus FLEXspace Typ 5230/5231
6 Operation | 6.6 Setting the response sensitivity
28/42