▶ Switch main device switch off.
▶ Clean the instrument and instrument with a moist, clean cotton cloth until no
residual soiling can be seen.
Automatic cleaning
Not applicable.
5.1.2
Cleaning the grip sleeve
Manual cleaning
▶ Brush off the grip sleeve under flowing, cold tap water until no visible contami‐
nation can be seen.
Automatic cleaning
CAUTION
Ultrasonic cleaning devices
Malfunctions from cleaning in ultrasonic cleaning devices
▶ Do not place the medical device in an ultrasonic cleaning device!
▶ Clean the grip sleeve in a thermodesinfector by Miele, for example.
▶ Use according to the manufacturer's instructions for use.
5.1.3
Cleaning the ZEG tips
Manual cleaning
▶ Brush off the ZEG tips under flowing, cold tap water until no visible contamination
can be seen.
Instructions for use PIEZOlux (0.571.5211 - 1.001.8522)
5 Preparation methods DIN EN ISO 17664 | 5.1 Cleaning
9/15