4 User interfaces
4.2
Patient positioning panel
1.
Carriage up/down slider
. The carriage can be driven up and down by sliding a finger on the carriage
up/down slider. The finger movement detection works best when using the entire finger tip or two
fingers side by side.
The speed of the carriage movement depends on the distance the finger is moved. For faster
carriage speed, start from the upper/lower edge of the slider.
2. HOME button
. Drives the rotating unit to HOME (Patient-In) and scan start positions. Pressing the
HOME button also wakes the device from Power save mode.
3.
Test mode button
. Disables radiation production of the device. You can also use the GUI to activate
the Test mode.
4. Patient positioning lights button
. Turns the patient positioning lights on/off.
5.
Exposure indicator
(YELLOW). Device is radiating.
6. Status indicator
(BLUE). Device is in error state or pending user action. Check the GUI for details.
7.
Ready indicator
(GREEN). Device is ready for imaging.
NOTICE!
If carriage up/down slider backlight is not illuminated, the carriage up/down slider is in
safe mode and cannot be used. Wait until the light comes back on or wipe the panel clean if it is
dirty and dry if it is wet.