7 Error codes and troubleshooting
169
7.6.8.2
Images too dark or too light (Wrong dose
generated)
•
Check the condition of the power supply by follow-
ing trouble shooting guide 7.6.2.3 for no X-rays.
•
Check that X-ray unit calibrations have been car-
ried out.
Preheat check
•
Set the X-ray unit to the 3D imaging mode (and us-
ing the s2terminal type patient), adjust the mA set-
tings from the GUI and then measure the
respective voltages from TP11 (Preheat refer-
ence).
mA check
•
Set the X-ray unit to the 3D imaging mode (and us-
ing the s2terminal type
patient
), adjust the mA set-
tings from the GUI and measure the respective
voltages from TP11 (mA reference).
If the FAIL LED comes on
replace R3810. If the LED
still comes on after re-
placing R3810, replace
the tubehead.
If voltages differ from those listed below, replace
EG200.
8 mA
1.54V
10 mA
1.59V
12.5 mA
1.65V
15 mA
1.70V
If the voltages do not differ, set the mA to 15 mA
and measure voltage from TP15 (preheat feed-
back) when an exposure is taken.
if voltage is not 0.58V then replace
R3810.
If this does not solve the problem
still, replace the tubehead.
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 18: ...1 Introduction 10...
Page 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Page 31: ...2 Unit description 23...
Page 32: ...2 Unit description 24...
Page 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Page 54: ...46...
Page 87: ...79...
Page 88: ...80...
Page 96: ...4 Firmware description 88...
Page 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Page 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Page 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Page 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Page 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Page 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Page 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Page 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Page 122: ...6 The s2terminal and service functions 114...
Page 192: ...8 Maintenance procedure 184...