7 Error codes and troubleshooting
126
7.4.7
Error 21 Calibration data receive
timeout
7.4.8
Error 22 Reading data from the 3D
sensor failed
CAUSE A
The main CPU did not receive calibration data from
the GUI PC in time.
SOLUTION A
• Timeout due to communication problems. See Er-
ror 100 for hints on how to trouble shoot cables and
Ethernet switch.
• Was the Cephalostat CPU recently replaced?
It may be loaded with invalid software for its pur-
pose and therefore interfering with other communi-
cation. Update the Cephalostat CPU with proper
software per instructions in pdf included in the em-
bedded software release package.
• Did the GUI PC start normally (or is the screen
dark and Error 21 reported in the s2terminal)?
If not, check that the GUI PC is powered properly.
Replace the GUI PC if needed. Note that the unit
then must be completely recalibrated. See 9 Cali-
bration and adjustment for further instructions.
CAUSE A
Bad contact in data cable connectors.
SOLUTION A
• Check the data cable (ribbon cable) is not dam-
aged and that it is properly connected both to the
3D sensor and to the R3220 main CPU board (con-
nector J3208). Reconnect or replace the cable if
needed.
• Clean the connectors with flux remover. Let dry
properly before reconnecting.
CAUSE B
Voltage level out of range or voltage missing.
SOLUTION B
Enable power for the 3D sensor using service com-
mand “panelpower conebeam 1”. Measure the dig-
ital voltage level from pin 1 and 2 on the sensor’s
power connector. The allowed range is 4.9…5.1 V.
If the voltage is missing, locate the point of failure
with the help of the schematic in 5.4.1 3D imaging
chain. Replace faulty parts.
CAUSE C
Faulty 3D sensor
SOLUTION C
Replace the 3D sensor and redo all 3D calibrations.
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 18: ...1 Introduction 10...
Page 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Page 31: ...2 Unit description 23...
Page 32: ...2 Unit description 24...
Page 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Page 54: ...46...
Page 87: ...79...
Page 88: ...80...
Page 96: ...4 Firmware description 88...
Page 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Page 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Page 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Page 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Page 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Page 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Page 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Page 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Page 122: ...6 The s2terminal and service functions 114...
Page 192: ...8 Maintenance procedure 184...