7 Cephalometric calibration and alignment
111
5. When the adjustments are done press the exposure but-
ton (no X-rays are generated) to acknowledge that the
check has been carried out.
7.2.7
Cephalostat pixel calibration
1. Attach secondary collimator and tubehead covers.
2. Rotate the ear holders into PA view position and move
them completely apart.
NOTICE! Turn nasion support up out of the way.
3. Select
CEPH Pix
Calibration program.
4. Press
Patient Positioning
.
5. Take an exposure.
6. This calibration should always be passed.
7.2.8
Cephalostat Quality Check (optional)
NOTICE! Rotate the ear holders into PA view position.
Turn nasion support up out of the way.
NOTICE! Not required in North America.
1. Attach the QC phantom to the cephalostat unit and en-
sure that it’s leveled from the spirit level.
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 8: ...iv rev...
Page 18: ...1 Introduction 10...
Page 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Page 29: ...2 Unit description 21...
Page 30: ...2 Unit description 22...
Page 38: ...4 PC requirements 30...
Page 54: ...5 Installing the unit 46 4 Fasten the covers...
Page 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Page 104: ...7 Cephalometric calibration and alignment 96...
Page 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Page 117: ...7 Cephalometric calibration and alignment 109...
Page 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Page 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Page 158: ...8 Special procedures 150...
Page 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Page 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Page 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...