Instructions for use MASTERmatic LUX M10 L - 1.009.3570
7 Reprocessing steps in accordance with ISO 17664 | 7.4 Drying
18 / 26
7.3.1 Manual disinfection - external
KaVo recommends the following products based on compatibility of the materials.
The microbiological efficacy must be ensured by the disinfectant manufacturer and
proven by an expert opinion.
▪ Mikrozid AF made by Schülke & Mayr (liquid or cloths)
▪ FD 322 made by Dürr
▪ CaviCide made by Metrex
Consumables required:
▪ Cloths for wiping off the medical device.
▶ Spray the disinfectant on a cloth, then wipe down the medical device and allow
the disinfectant to act according to the instructions of the disinfectant manufactur‐
er.
▶ Follow the instructions for use of the disinfectant.
7.3.2 Manual disinfection - internal
The efficacy of manual internal disinfection must be demonstrated by the manufactur‐
er of the disinfection agent. With KaVo products, use only disinfection agents that
have been released by KaVo, with respect to the compatibility of materials
(e.g. WL-cid / made by ALPRO).
▶ Blow off with compressed air until no water drops are visible.
▶ Immediately after internal disinfection, lubricate the KaVo medical device immedi‐
ately with care agents from the KaVo care system.
7.3.3 Machine disinfection - external and internal
KaVo recommends washer disinfectors in compliance with EN ISO 15883-1, which
are operated using alkaline cleaning agents.
▶ For programme settings as well as cleansers and disinfectants to be used, please
refer to the Instructions for Use of the washer disinfector.
▶ In order to prevent negative effects on the medical device, make sure that the in‐
terior and the exterior of the medical device are dry after completion of the cycle,
and then grease it immediately with servicing agents from the KaVo care system.
7.4 Drying
Manual Drying
▶ Blow off the outside and inside with compressed air until water drops are no lon‐
ger visible.
Automatic Drying
The drying procedure is usually a part of the cleaning program of the thermodisinfec‐
tor.
▶ Follow the instructions for use of the thermodisinfector.