3 Product description | 3.6 Technical data and requirements
20 / 110
3.6 Technical data and requirements
Photobiological safety (IEC 62471)
Risk group 1 (low risk) The products
are safe under the majority of meth-
ods of application, except for the
case of prolonged exposition with
potentially direct eye exposure.
Electrical system for mounting the device
Input voltage
24 V AC
Frequency
50/60 Hertz
Power consumption
max. 35 VA
Ceiling mounting and electrical system
Input voltage
100 – 240 V AC
Frequency
50/60 Hertz
Power consumption
max. 50 VA
Electrical lead
3x1.5 mm
2
in accordance with DIN
VDE 0100-710
Free end at assembly point
1000 mm
Customer-provided fuse protection
Automat C 16 or screw-plug fuse 10
A. Can be connected to the auto-
matic unit of the dental treatment
centre
Main fuse of the terminal
T 500mA H / 250 V
Colour reproduction and illumination strength
Colour temperature normal light
ca 4.000 to 6.000 Kelvin; preset
5.500 Kelvin (daylight quality)
Colour rendering index CRI
>93 at 5,500 Kelvin Δc < 0.005
Normal light
approx. 20,000 to 40,000 lux; preset
30,000 lux in accordance with
ISO 9680:2015
COMPOsave (compatible mode for
light-activatable restoration materi-
als)
approx. 8,000 to 15,000 Lux in ac-
cordance with ISO 9680:2015
Dimmed light
approx. 8,000 to 15,000 Lux / ap-
prox. 4,000 K
Typical distribution of illumination in-
tensity
blue 75%, green 50%, yellow 10%
of the illumination intensity