Technician's instructions KEY Laser III 1243
2 Measuring and configuration
|
2.2 Optical system
▶ Use a cotton bud with a wooden stem to prevent the energy measurement unit
mirror from slipping.
▶ Place a sheet of lens cleaning paper on the energy measurement unit mirror
and, without applying any pressure, wipe the surface of the mirror.
Wipe at a speed whereby the acetone evaporates before it reaches the end of
the lens cleaning paper.
▶ Clean the second surface of the energy measurement unit mirror and the two
surfaces of the pilot input unit mirror in the same way.
Install the deflection mirror
Note
The < character on the lateral surface of the mirror points toward the coating.
The installation direction of the energy measurement mirror ③ and the pilot
input ④: Coating in the direction of the Er:YAG laser illumination.
Do not confuse the two mirrors. Install them in the correct holder with the correct
cover according to the drawing.
99/176
Summary of Contents for KEY Laser III 1243
Page 1: ...Technician s instructions KEY Laser III 1243 Always be on the safe side ...
Page 179: ......