▶ Pull off the handpiece by twisting it axially.
▶ To prevent soiling the hose coupling window ② wipe off the laser hose coupling
with a clean, dry cloth from back to front.
See also: Section A 2.3 Handpiece instructions.
Note
Windows are coated. Do not wipe unnecessarily.
▶ Check if the window ② is clean and dry, and if the O-rings ③ are undamaged.
Note
If a handpiece is not coupled directly afterwards, the laser hose coupling must be
provided with the cap ① and the handpiece with the cap ④. These two caps prevent
the penetration of dirt into the laser hose coupling and the handpiece.
▶ Dry the inside of the handpiece coupling.
4.4
Switching on
Note
Make sure that the emergency-off switch is not pressed. This can be released or
its condition can be checked by turning the red knob to the right. If it was pressed,
it will spring out.
Instructions for use KEY Laser 3+ 1343 KEY Laser III 1243 upgraded
4 Operation | 4.4 Switching on
34/76