6.7.1 Care products and systems - Servicing: Care with KaVo Spray
KaVo recommends servicing the project twice daily (at noon and in the
evening after hours), after each time the machine is cleaned, and before
each sterilisation.
▶
Remove the cutter or grinder and close the chuck.
▶
Cover the product with the Cleanpac bag.
▶
Place the product on the cannula, and press the spray button for one
second.
Preparation methods according to ISO 17664
47