Operating Instruction GENTLEray 982 Bleaching
3 Setup methods according to DIN EN ISO 17664
|
3.3 Sterilisation in a steriliser in compliance with DIN EN
13060
3.3 Sterilisation in a steriliser in compliance with DIN EN 13060
Note
All handpieces must be prepared. Preparation (cleaning and sterilisation) should
be carried out directly after each use.
▶ Always cover the lens with the dust cap ②.
▶ Screw the bleaching sleeve ① off the cap ③ and remove.
▶ Put the bleaching sleeve ① in a bag.
Note
The lens connector including the lens and bleaching fibre may not be sterilised.
They must be cleaned or disinfected.
▶ The bleaching sleeve can be sterilised in a fractionated initial vacuum at 134°C
± 1°C, 3.04 bar, for 4 minutes (sterilisable up to max. 138 °C).
Note
The bleaching sleeve can be steam sterilised (up to 135°C, 2.1 bar, 10 minutes).
CAUTION
The bleaching fibre should not be sterilised
This can damage the bleaching fibre.
▶ Clean and disinfect the bleaching fibre when it is becomes dirty.
13/27