![KaVo EXPERTmatic LUX E25 L Instructions For Use Manual Download Page 11](http://html1.mh-extra.com/html/kavo/expertmatic-lux-e25-l/expertmatic-lux-e25-l_instructions-for-use-manual_3818149011.webp)
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551
3 Description of the product | 3.2 Technical Specifications
11 / 28
▪ to prevent contamination from the product
3.2 Technical Specifications
Drive speed
max. 40,000 rpm
Speed transmission
1 : 5
Identification
1 red ring
Spray water pressure
0.8 to 2.0 bar (12 to 29 psi)
Spray air pressure
1.0 to 2.0 bar (15 to 29 psi)
Spray air flow rate
min. 1.5 Nl/min (at 2 bar)
Cooling air flow
5.5 to 9.5 Nl/min
Pushbutton chuck
Ø 1.6 mm
Insert
Dental burs and diamond grinders in
accordance with DIN EN ISO 1797
type 3
Can be attached to
All INTRA (LUX) motors and motors
with a connector in accordance with
DIN EN ISO 3964
3.3 Transportation and storage conditions
CAUTION
Startup after refrigerated storage.
Malfunction.
4
Prior to startup, strongly refrigerated products must be allowed to warm up
to a temperature of 20 °C to 25 °C (68 °F to 77 °F).
Temperature: -29 °C to +50 °C (-20 °F to +122 °F)
Relative humidity: 5% RH to 85% RH absence of condensation
Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi)
Protect from moisture (Keep dry)