Instructions for use EXPERTmatic LUX contra-angle E25 L - 1.007.5550 EXPERTmatic contra-angle E25 C - 1.007.5551
4 First use | 4.1 Check the amount of water
4 First use
WARNING
Hazard from non-sterile products.
Infection hazard for care provider and patient.
▶ Before first use and after each use, prepare and sterilise the medical device and
accessories accordingly.
CAUTION
Damage from soiled and moist cooling air.
Contaminated and moist cooling air can cause malfunctions and lead to premature
bearing wear.
▶ Make sure that the supply of cooling air is dry, clean and uncontaminated accord‐
ing to ISO 7494-2.
4.1 Check the amount of water
CAUTION
Overheating of the tooth due to lack of cooling water.
Thermal damage to the dental pulp.
▶ Adjust the water amount for the spray cooling to a minimum of 50 cm
3
/min
CAUTION
Hazard from insufficient spray water.
Insufficient spray water can cause the medical device to overheat and damage the
tooth.
▶ Check the spray water channels and clean the spray nozzles with the nozzle nee‐
dle Mat. no. 0.410.0921 if necessary.