Instructions for use EXPERTmatic E31 C – 1.008.4998
7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.4 Care products and systems - Servicing
21 / 28
NOTE
Please comply with the instructions for use of the washer disinfector.
4
In order not to affect the KaVo medical device, make sure that the product
is dry on the inside and outside after completion of the cycle.
4
Remove any residual liquids with KaVoDRYspray.
Also refer to:
4
Immediately after drying, lubricate the KaVo medical device with care
agents from the KaVo care system.
7.4 Care products and systems - Servicing
CAUTION
Improper service and care.
Risk of injury.
4
Service regularly with suitable agents.
NOTE
Remove the servicing dental bur.
NOTE
KaVo guarantees the proper function of KaVo products only if the care
products listed by KaVo as accessories are used, since these were tested for
proper use on our products.
7.4.1 Servicing with KaVo Spray
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation, but no
later than after 30 minutes of operation.
4
Remove the dental bur from the medical device.
4
Cover the medical device with the KaVo Cleanpac bag, and place it on the
corresponding care adapter.
4
Press the spray key for 1 to 2 seconds.
7.4.2 Servicing with KaVo QUATTROcare PLUS
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation, but no
later than after 30 minutes of operation.
4
Remove the dental bur from the medical device.
4
Service the device in the QUATTROcare PLUS.
Also refer to:
Instructions for use KaVo QUATTROcare PLUS
7.5 Packaging
NOTE
The sterile goods package must be large enough to accommodate the product
without stretching the packaging. The quality and use of the packaging of the
items to be sterilised must meet the applicable standards and be appropriate
for the sterilisation process!
4
Seal the medical device separately in a sterile pack.